Distek 2100c Manual

2100c

Distek 2100b Manual

The Distek 2100C Dissolution System with Agilent 8453 UV-Vis Spectroscopy System provides automatic control over “infinity RPM” at the end of test, comprehensive method reporting, remote programming and control, validation, and documentation. The Agilent 8453 UV-Vis Spectroscopy is a microprocessor control, with collimating optics that can be controlled with either a HP or non-HP computer.
Specifications for 2100C Dissolution System:
- Electrical Power: (Operating voltage pre-set at factory) 115V 50/60Hz 0.5A or 220V 50/60Hz 0.5A
- Dissolution Vessels: 7 built-in vessel positions
- Volume: 500mL to 1000mL
- Drive Positions: 6 positions standard
- RPM Control Range: 25-250 RPM, digitally controlled, closed loop
- Resolution: 0.1 RPM
- Accuracy: +/- 0.5 RPM
- Display: 0.55 inch (14mm) four-digit LED
- Motor: High Torque, Permanent Magnet
- Shaft Wobble: less than 0.010” (0.254mm) TIR (total indicator reading)
- Bath Temperature Control: TCS-0200C Thermocirculator
- Display Resolution: 0.1° C
- Accuracy: +/- 0.25° C
- Program Modes: Manual, Automatic, External
- Construction Materials: Aluminum, Stainless Steel, Engineered Plastics
- Dimensions (W x H x D): 25” x 39” x 19¼” in. (63.5 x 99 x 49cm Deep)
- Net Weight (approx.): 140 lb. (63.5kg) (water bath full)
Specifications for Agilent 8453 UV-Vis Spectroscopy System:
- Line voltage: 90–264 V AC, Wide-ranging capability
- Line frequency: 47–63 Hz
- Power consumption: 220 VA, Maximum
- Ambient operating temperature: 0–55 °C (32–131 °F)
- Ambient non-operating temperature: -40–70 °C (-4–158 °F)
- Humidity: <95%, at 25–40 °C (77–104 °F), Non-condensing
- Operating altitude: Up to 2000 m (6,500 ft)
- Non-operating altitude: Up to 4600 m (14,950 ft), For storing the instrument
- Safety standards: IEC, CSA, UL Installation Category II, Polution Degree 2
- Dimensions ( W x D xH): 34.4 x 56.0 x 18.5 cm (13.5 x 22.0 x 7.3 inches) wide
- Weight 16.5 kg (36.3 lbs)

Distek 2100c Manual Instructions

The Distek 2100C Dissolution System with Agilent 8453 UV-Vis Spectroscopy System provides automatic control over “infinity RPM” at the end of test, comprehensive method reporting, remote programming and control, validation, and documentation. The Agilent 8453 UV-Vis Spectroscopy is a microprocessor control, with collimating optics that can. Download KEITHLEY 2100 6,5-DIGIT VFD DMM 2008 CALIBRATION SM service manual & repair info for electronics experts.